52-3.2 Sterilization of instruments.
All instruments, including collection containers, used for both sterile and clean, nonsterile retrieval of tissues shall be either licensed by the FDA for single use or be sterilized prior to each use. Heat sterilization shall be by autoclaving at 121.5 degrees Celsius for 15 minutes after the chamber of the autoclave has been evacuated and has reached that temperature, or by dry heat for two hours at 170 degrees Celsius, or by such other procedures as may be approved by the department upon submission of evidence that such other procedures are at least equally effective.
10 CRR-NY 52-3.2
Current through May 31, 2021
End of Document

IMPORTANT NOTE REGARDING CONTENT CURRENCY: JULY 31, 2023, is the date of the most recently produced official NYCRR supplement covering this rule section. For later updates to this section, if any, please: consult editions of the NYS Register published after this date; or contact the NYS Department of State Division of Admisnistrative Rules at [email protected]. See Help for additional information on the currency of this unofficial version of the NYS Rules.