§ 30316.61. Instruments Used by Mammography Medical Physicists.

17 CA ADC § 30316.61Barclays Official California Code of Regulations

Barclays California Code of Regulations
Title 17. Public Health
Division 1. State Department of Health Services (Refs & Annos)
Chapter 5. Sanitation (Environmental) (Refs & Annos)
Subchapter 4. Radiation
Group 3. Standards for Protection Against Radiation
Article 4.5. Requirements for the Use of X-Ray in Mammography
17 CCR § 30316.61
§ 30316.61. Instruments Used by Mammography Medical Physicists.
(a) Instruments used by mammography medical physicists to measure the air kerma or air kerma rate from a radiation machine used for mammography shall be calibrated at least once every two years and each time the instrument is repaired. The instrument calibration shall be traceable to a national standard and calibrated with an accuracy of plus or minus 6.0% (95% confidence level) in the mammography energy range.

Credits

Note: Authority cited: Sections 114975, 115000, 115060 and 131200, Health and Safety Code. Reference: Sections 115060, 115100, 131050 and 131051, Health and Safety Code.
History
1. New section filed 2-10-2003 as an emergency; operative 2-10-2003 (Register 2003, No. 7). A Certificate of Compliance must be transmitted to OAL by 6-10-2003 or emergency language will be repealed by operation of law on the following day.
2. Certificate of Compliance as to 2-10-2003 order transmitted to OAL 6-6-2003 and filed 7-18-2003 (Register 2003, No. 29).
3. Amendment of section heading, section and Note filed 12-18-2019; operative 7-1-2020 pursuant to Government Code section 11343.4(b)(2) (Register 2019, No. 51).