§ 30316.20. Quality Assurance Testing.
17 CA ADC § 30316.20Barclays Official California Code of Regulations
Barclays California Code of Regulations
Title 17. Public Health
Division 1. State Department of Health Services (Refs & Annos)
Chapter 5. Sanitation (Environmental) (Refs & Annos)
Subchapter 4. Radiation
Group 3. Standards for Protection Against Radiation
Article 4.5. Requirements for the Use of X-Ray in Mammography
17 CCR § 30316.20
§ 30316.20. Quality Assurance Testing.
(a) Each facility using screen-film systems for mammography shall adjust and maintain each processor used to develop mammograms so that the manufacturer's technical development specifications for the mammography film in use are met. Each day mammograms are processed and before processing mammograms the facility shall determine that the processor used to develop mammograms, using the mammography film of the type used clinically, meets the following:
(b) Each facility using screen-film systems for mammography shall, each week in which mammography is performed, produce an image of a phantom. The test shall be performed prior to mammography on the day the test is performed using film processors that may be used to develop mammograms. Before exposing the phantom, an acrylic disc measuring one centimeter in diameter and four millimeters thick shall be placed on the phantom in the image area so it will not obscure details in the phantom and where it cannot cast a shadow on any portion of the AEC detector. The phantom shall then be exposed using the mammography film of the type used clinically and the techniques used for clinical images of a standard breast. The resulting phantom image shall meet the following:
(2) Perform a repeat analysis on mammograms repeated or rejected. If the total repeat or reject rate changes from the previously determined rate by more than 2.0% of the total films included in the analysis, the reason(s) for the change shall be determined by the facility. Any corrective actions shall be documented and the results of those corrective actions shall be assessed. Test films, cleared films, or film processed as a result of exposure of a film bin shall not be included in the count for repeat analysis but shall be counted to determine reject rate changes and may be disposed of following completion of the analysis.
(1) When conducting screen-film mammography, determine that the optical density attributable to darkroom fog does not exceed 0.05, by performing a test which uses mammography film of the type used clinically in the facility in which the film is exposed such that the film has a mid-density of no less than 1.2 OD, and is exposed to typical darkroom conditions for two minutes while such film, with one-half of the film covered, is placed on the counter top, emulsion side up. If the darkroom has a safelight used for mammography film, it shall be on during this test;
(3) Determine that the X-ray system is able to compress the breast with a force of at least 25 pounds and maintain this compression for at least 15 seconds, except that for systems with automatic compression, the maximum force applied without manual assistance shall be greater than 25 pounds and shall not exceed 45 pounds.
(1) The AEC can maintain film optical density within plus or minus 0.15 of the average of the optical densities measured using homogeneous breast-tissue equivalent material thicknesses of 2, 4, and 6 centimeters (cm) and the kilovoltage peak (kVp) is varied appropriately for such thicknesses over the kVp range used clinically in the facility. Each image of the homogenous breast-tissue equivalent material shall have an OD of at least 1.20;
(A) Each mammography system used for mammography, in combination with the mammography screen-film combination used in the facility, shall provide a minimum resolution of 11 cycles/millimeter (mm) (line-pairs/mm) when a high contrast resolution bar test pattern is oriented with the bars perpendicular to the anode-cathode axis, and a minimum resolution of 13 line-pairs/mm when the bars are parallel to that axis;
(E) Test kVp shall be set at the value used clinically by the facility for a standard breast and shall be performed in the AEC mode, if available. If necessary, a suitable absorber may be placed in the beam to increase exposure times. The screen-film cassette combination used by the facility shall be used to test for this requirement and shall be placed in the normal location used for clinical procedures.
(6) The half-value layer (HVL) of the useful beam is not less than the value specified in Table 2 of section 30308(a)(3) for the minimum HVL. These values, extrapolated to the mammographic range, are shown as follows, except that values not shown in Table 2 may be determined by linear interpolation or extrapolation:
Table 2
X-ray Tube Voltage (kilovolt peak) and Minimum HVL |
---|
Designed Operating Range (kV) | Measured Operating Voltage (kV) | Minimum HVL (mm of aluminum) |
---|---|---|
Below 50 | 20 | 0.20 |
25 | 0.25 | |
30 | 0.30 |
(8) By performance of a test using the techniques the facility uses clinically for a standard breast, the average glandular dose delivered by screen-film systems during a single craniocaudal view of an FDA accepted phantom simulating a standard breast does not exceed 3.0 milligray (mGy) (0.3 rad) per exposure;
(B) Provide that if a light field that passes through the X-ray beam limitation device is used, it is aligned with the X-ray field so that the total of any misalignment of the edges of the light field and the X-ray field along either the length or the width of the visually defined field at the plane of the breast support surface shall not exceed 2.0% of the SID; and
(C) The chest wall edge of the compression paddle does not extend beyond the chest wall edge of the image receptor by more than 1.0% of the SID when tested with the compression paddle placed above the breast support surface at a distance equivalent to standard breast thickness. The shadow of the vertical edge of the compression paddle shall not be visible on the image.
(11) During the uniformity of screen speed test specified in paragraph (10), system artifacts are evaluated with a high-grade, defect-free sheet of homogeneous material large enough to cover the mammography cassette and performed on all cassette sizes used in the facility using a grid appropriate for the cassette size being tested. The optical density of the test films shall be no less than 1.4;
(13) Each machine produces, over 3.0 seconds, a minimum output of 7.0 mGy air kerma per second (800 milliroentgen (mR) per second) when operating at 28 kVp in the standard mammography (molybdenum/molybdenum) mode at any SID where the system is designed to operate and when measured by a detector with its center located 4.5 cm above the breast support surface with the compression paddle in place between the source and the detector;
(16) For systems with image receptor modalities other than screen-film, the quality assurance program meets the quality assurance program recommended by the image receptor manufacturer, except that the maximum allowable dose shall not exceed the maximum allowable dose for screen-film systems specified in paragraph (8).
(g) After completion of the tests specified in subsections (a) through (f), the test results shall be compared to the corresponding specified action limits. For non-screen-film modalities, the test results shall be compared to the manufacturer's recommended action limits. If the test results fall outside of the action limits, the source of the problem shall be identified and corrective actions shall be taken:
Credits
Note: Authority cited: Sections 114975, 115000, 115060 and 131200, Health and Safety Code. Reference: Sections 115060, 115100, 115115, 131050 and 131051, Health and Safety Code.
History
1. New section filed 2-10-2003 as an emergency; operative 2-10-2003 (Register 2003, No. 7). A Certificate of Compliance must be transmitted to OAL by 6-10-2003 or emergency language will be repealed by operation of law on the following day.
2. Certificate of Compliance as to 2-10-2003 order, including amendment of subsections (b)-(b)(2), (c), (e)(2)-(3), (e)(6)(A), (e)(13) and (f)(2)-(3), transmitted to OAL 6-6-2003 and filed 7-18-2003 (Register 2003, No. 29).
3. Amendment of section and Note filed 12-18-2019; operative 7-1-2020 pursuant to Government Code section 11343.4(b)(2) (Register 2019, No. 51).
This database is current through 4/26/24 Register 2024, No. 17.
Cal. Admin. Code tit. 17, § 30316.20, 17 CA ADC § 30316.20
End of Document |