Home Table of Contents

Appendix E--Attachment B

8 CA ADC § 5207 App. E Attach. BBarclays Official California Code of Regulations

Barclays California Code of Regulations
Title 8. Industrial Relations
Division 1. Department of Industrial Relations
Chapter 4. Division of Industrial Safety
Subchapter 7. General Industry Safety Orders
Group 16. Control of Hazardous Substances
Article 110. Regulated Carcinogens
8 CCR § 5207 App. E Attach. B
Appendix E--Attachment B
Instrumental Parameters for HGA Analysis
Atomic Absorption Spectrophotometer (Perkin-Elmer Model 5100)
Signal Type:
Zeeman AA
Slitwidth:
0.7 nm
Wavelength:
228.8 nm
Measurement:
Peak Area
Integration Time:
6.0 sec
BOC Time:
5 sec
BOC = Background Offset Correction
Zeeman Graphite Furnace (Perkin-Elmer Model HGA-600)
Step
Ramp Time (sec)
Hold Time (sec)
Temp. (°C)
Argon Flow (mL/min)
Read (sec)
1)Predry
5
10
90
300
--
2)Dry
30
10
140
300
--
3)Char
10
20
900
300
--
4)Cool Down
1
8
30
300
--
5)Atomize
0
5
1600
0
-1
6)Burnout
1
8
2500
300
--
This database is current through 6/21/24 Register 2024, No. 25.