Home Table of Contents

005.28.52-6.0. Defibrillator Preventative Maintenance/Repair

AR ADC 005.28.52-6.0Arkansas Administrative CodeEffective: May 30, 2022

West's Arkansas Administrative Code
Title 005. Department of Education
Division 28. Division of Elementary and Secondary Education
Rule 52. Division of Elementary and Secondary Education Rules Governing School-Based Automated External Defibrillator (AED) Devices and Cardiopulmonary Resuscitation (CPR) Programs in Arkansas Public Schools
Effective: May 30, 2022
Ark. Admin. Code 005.28.52-6.0
005.28.52-6.0. Defibrillator Preventative Maintenance/Repair
6.01 Each school district shall designate appropriate personnel to be responsible for the maintenance of the AED(s).
6.02 All components of the AED and integrated data recording system shall be inspected by a qualified service technician at least one (1) time per calendar year or as recommended by the manufacturer to ensure:
6.02.1 The equipment meets original manufacturer's specifications;
6.02.2 The equipment maintains the currently approved treatment protocols based on the current American Heart Association scientific guidelines, standards, and recommendations for the use of the AED.
6.03.2 The battery of the AED shall be maintained and replaced in accordance with manufacturer's specifications.
6.03.3 All maintenance and repairs shall be performed by a qualified service technician recognized by the manufacturer.
6.03.4 Written records shall be maintained for all maintenance, repairs, and inspections performed on all components for mandated annual state reporting purposes.

Credits

Adopted May 30, 2022.
Current with amendments received through February 15, 2024. Some sections may be more current, see credit for details.
Ark. Admin. Code 005.28.52-6.0, AR ADC 005.28.52-6.0
End of Document